Brooks Instrument GC Series Compact Gas Mass Flow Controllers
Product Overview
Brooks Instrument GC Series (GC50/GC51) offer supercompact gas flow control, specifically designed for benchtop laboratory systems and cost-sensitive OEM platforms. By leveraging advanced MEMS (Micro-Electro-Mechanical Systems) technology, the GC Series delivers high-performance thermal mass flow control in an incredibly small footprint. This ultra-compact design allows for seamless integration into space-limited environments such as bioreactors, analytical equipment, and medical analyzers without compromising on precision or stability.
Engineered for versatility, the GC Series features a wide 100:1 turndown ratio and a flow range up to 5000 sccm, allowing a single device to cover a broad array of operating conditions. Users benefit from simplified system designs and reduced spare parts inventory thanks to the in-situ verification and recalibration capabilities provided by Brooks’ BCAT software. With multiple connectivity options including Modbus RTU (RS-485) and various analog interfaces, the GC Series is a flexible, plug-and-play solution for modern laboratory and industrial gas delivery systems.
Features and Benefits
Take a look at some of the features and benefits of the GF40 Series
In-Situ Verification & Recalibration: Utilize Brooks Instrument BCAT software to verify accuracy and recalibrate units without removing them from the process line, drastically reducing maintenance downtime.
Inventory Consolidation: The multi-gas/multi-range capability allows a single device to be reconfigured for different setups, reducing the need to stock multiple spare part SKUs.
“Matchbox” Form Factor: The ultra-compact footprint allows for easy integration into dense OEM manifolds or crowded benchtop workstations where traditional MFCs won’t fit.
Simplified System Integration: Dual-mode communication (Analog and Digital) ensures the device “drops in” to legacy systems or modern PLC-controlled environments without needing expensive adapters.
Enhanced Process Integrity: Superior valve shut-off minimizes “leak-by” during idle periods, protecting sensitive samples in bioreactors or analytical equipment from unwanted gas exposure.
Global Regulatory Compliance: Fully certified for safe area use and global standards, making it the ideal choice for OEMs exporting equipment to international markets.
Specifications
For full specifications, including electrical, diagnostics, and compliance information, download the product specification sheet below.
Full Scale Flow Range
210 sccm up to 5,000 sccm (N2 equivalent)
Control Accuracy
± 1.5% of Rate of S.P. ± 0.5 of F.S.
Turndown Ratio
100:1 (Provides wide control range with a single device)
Repeatability
± 1.5% of Set Point
Communication Protocols
Modbus RTU (via RS-485), 0-5V, 0-10V, 4-20mA
Physical Dimensions
22mm x 40mm x 60mm
Operating Pressure
Up to 150 psi (10.3 bar) maximum