Understanding & Customizing Your GPI “09” Digital Display

Bill Michie | November 12th, 2014

 

Calibrating your GPI “09”Series Electronic display using the older “Display/Dispense”

In the fourth and final installment of this series, we’ll discuss the procedures for calibrating your GPI “09”Series Electronic display using the older “Display/Dispense” method. As mentioned in an earlier post, this field calibration procedure has been programmed into the 09 Series Display/Flow Computer, but is disabled by default and must be enabled using the custom programming instructions in Part 3 in order for you to access it. Click Here to view instructions for accessing/changing custom configuration settings.

Once “Display/Dispense” has been enabled, use the following instructions to field calibrate your GPI Flow Meter via this method.

  1. While in the Factory Calibration mode (FAC will be visible in the upper right hand corner of your display), hold down both the Calibrate and Display buttons for approximately 3 second until you see “FLdCAL” on the display, then let go of the buttons. Note: dd000.0 will appear on the screen
  2. For the most accurate results, it’s recommended that you use a uniformly dependable and accurate calibration container. Your flow meter should be installed and purged of any air prior to beginning the test.The first step is to flow a known amount of fluid (by weight or volume) through the flow meter while in the Field Calibration mode. It’s highly recommended that you dispense fluids at a flow rate which best simulates your actual operating conditions. Avoid “dribbling” more fluid or repeatedly starting and stopping the flow. This can result in less accurate calibrations.
  3. The flow meter display will count up as the fluid passes through the meter. Compare the final reading on the display to the known amount. If the totals match, the meter is correct as-is and will not require any additional field calibration. Skip ahead to Step #5.
  4. If the final reading on the display does not match the known amount, change the amount on the display to match using the following procedure:
  5. Begin by pushing the “Display” button. The first digit to the left will begin to blink. If it needs to be changed, use the “Calibrate” button to scroll through the available digits until correct.
  6. Push and release the “Diplay” button again to move to the next digit to the right. Using the “Calibrate” button, repeat the procedure as needed until the screen reading matches the known amount.
  7. Once the display reading has been entered correctly and/or if no adjustment was necessary, hold down both buttons (Calibrate and Display) until “CALEnd” appears on the screen. Field calibration is now complete.

To switch back to your original factory calibration, hold down both buttons again until FAcCAL appears on the screen.

Latest Cross Company Updates

From beer to soft drinks, high-quality, food-grade hoses are necessary to enable a clean, sanitary, and safe fluid transfer during the production process. See how a solution from Parker can help improve your process.
October 1, 2026
We’ve added solid metal conductivity and torque multipliers to our 17025 scope in Atlanta! See what else we’ve added!
September 28, 2026
Cross Logo
When dealing with older machined components like plastic injection molds or stamping dies, maintaining and restoring tools without existing CAD files can be a massive hurdle.
September 10, 2026
It is obvious that for the process to work efficiently and without problems, it is necessary to choose the right equipment in the right places for the process. This once again proved to be true, when Michigan Sugar was facing problems
September 8, 2026
Gases like hydrogen, helium and oxygen are in higher demand than ever before. This means the equipment used in these units and applications must maintain its reliability and provide longevity within the process.
September 8, 2026

Contact our Team

"*" indicates required fields

This field is for validation purposes and should be left unchanged.
What can we help you with today?*

Where Can We Send Your Information?

Name*
Company Location*
Form Type

Hang Tight! We're Searching... Searching... Searching...

We’re looking through thousands of pages to find the most relevant information.

In the meantime, enjoy these fun facts…

Did you know… Cross Company is an ESOP (Employee Stock Ownership Plan). Our ESOP started in 1979 and as of 2006, we are 100% employee-owned! Learn more about our ESOP and how that benefits both team members and our customers.
Did you know... the precision measurement group at Cross was founded in 1939 by our current CEO's grandfather, Jim King. That's a whole lot of calibration!
Did you know... A fingerprint weighs about 50 micrograms. We know, we weighed it! The residue left from a finger can actually make a difference in weight results which is why we wear gloves when we calibrate weights. For reference, a sheet of paper is about 4.5 grams, that’s 4.5 million micrograms.
Did you know… Cross Company has grown significantly since our start in 1954. Over the years we've acquired 26 companies! Today, our five groups have expertise in everything from industrial automation to precision measurement, and industry knowledge going all the way back to 1939.